Section- A
Sensor Head Alignment and Calibration
Mercury II
6000 Models – Setup using Software
B2.4.3
Calibrating the Encoder
Once the SmartPrecision II Software is open,
the encoder can be aligned and calibrated
using the Calibrate and Align tab. Click on the
Calibrate and Align tab and perform sensor
alignment using the Align Mode “Turn On”
and “Turn Off” buttons. Turn on Align Mode,
adjust the sensor until the maximum signal
strength is achieved, tighten the sensor
mounting screws, check for index indication
(using the status display at the top right of the
screen), and turn off Align Mode.
Next, perform setup by checking the Calibrate
GOP, Calibrate Index, Set Left Limit, and Set
Right Limit boxes, as shown.
Ensure that the calibration type is correctly selected for your encoder and
press Start. Follow the steps in the Calibration Status Box to complete setup of
the MII 6000 encoder.